Moisture & Dew Point Measurement Series
CONG Prima-10

CONG Prima-10 Online Water & Hydrocarbon Dew Point Analyzer

Two-component online chilled-mirror system for high-accuracy water-dew-point and hydrocarbon-condensation measurement in natural gas.

Brand: Vympel Manufactured by: LLC Scientific Production Association "Vympel", Russia Supply & support: Blue Dragon Technology
Product Overview

Purpose-built measurement, clearly presented.

The CONG Prima-10 is a two-component online analyzer system comprising an explosion-protected Dew Point Transducer and a Central Control Unit installed in a safe area. One chilled-mirror measurement cell can register water dew point, hydrocarbon condensation temperature, or both values alternately.

The system is available with a gas-preparation arrangement or a direct-pipeline module. Blue Dragon Technology reviews the measurement ranges and accuracy, line and measurement pressure, gas composition, contamination risks, supplemental cooling, sample-system architecture, installation areas, Central Control Unit location, input and output requirements, and commissioning method before configuration.

Distinctive Capabilities

  • Laboratory-quality chilled-mirror measurement in an online field system
  • Water and hydrocarbon condensation measurement with one transducer cell
  • Central Control Unit for operation, diagnostics, and plant data interfaces
  • Sample-system and direct-pipeline installation configurations
  • Smart mirror-cleaning function with no routine analytical consumables
Performance & Technical Data

The measurement method and the values that matter.

How it measures

Measurement Principle

The Dew Point Transducer cools a mirror while a laser-based interferometric optical system monitors the surface for condensation. The mirror temperature at the detected onset of water or hydrocarbon condensation is registered by the transducer. The Central Control Unit provides operational control, diagnostics, adjustment, and digital and analogue plant interfaces.

Measurement ParametersWater dew point and hydrocarbon condensation temperature
Analyzer TypeTwo-component automatic online chilled-mirror system
Measurement PrincipleLaser-based interferometric condensation measurement
Water Dew Point Range-30 to +30 °C; -50 to +10 °C with supplemental cooling
Hydrocarbon Condensation Range-30 to +30 °C; -30 to +10 °C
Water Dew Point Accuracy±0.5 °C; ±0.25 °C within -30 to +30 °C

Complete Technical Data

Hydrocarbon Condensation Accuracy
±0.5 °C
Measurement Frequency
4-14 cycles/hour maximum
Sample Gas Flow
0.5-2.0 Nl/min
Maximum Gas Sample Pressure
≤160 bar
Ambient Temperature
-40 to +40 °C
Explosion Protection
ATEX II 2G Ex d IIA T5 Gb
Ingress Protection
IP54
Central Control Unit Outputs
RS-485 Modbus RTU; RS-232 Modbus RTU; 2 × 4-20 mA
Central Control Unit Inputs
4-20 mA thermometer; 4-20 mA pressure transmitter
Supplemental Cooling
Required for water dew-point measurements below -30 °C
System Power Supply
Configuration dependent; confirm Central Control Unit and accessory supply at order
Dimensions
Central Control Unit: 483 × 320 × 128 mm; transducer: 207 × 112 × 235 mm
Weight
Central Control Unit: 6.5 kg; transducer: 6.5 kg
Industrial Use

Applications and industries.

Typical Applications

Natural Gas Quality MonitoringWater Dew Point MeasurementHydrocarbon Condensation Temperature MeasurementGas Dehydration MonitoringPipeline Gas Quality MonitoringGas Moisture Concentration MeasurementTrace Moisture Measurement in Gases

Industries Served

Chemical ProcessingNatural Gas & LNGOil & GasPipelines & MidstreamSulfur Recovery & Gas Treating
Engineering & Integration

Built for specification, installation, and long-term operation.

The CONG Prima-10 should be engineered as a complete online measurement system. The transducer, safe-area Central Control Unit, sample or pipeline interface, cooling, pressure and temperature inputs, utilities, communications, and maintenance access must be defined together.

System TypeTwo-component online analyzer
Primary ComponentsDew Point Transducer and Central Control Unit
InstallationGas-preparation system or direct-pipeline module
Project FocusSample conditioning, cooling, pressure and temperature inputs

System Configuration

  • OptionalGas-preparation system, supplemental cooling, pressure and temperature transmitters, direct-pipeline module, filtration, and zero-emission sample arrangement
  • Project EngineeredExtraction, pressure and flow control, enclosure or shelter, heating, cooling, drainage, sample return, cabling, safe-area interfaces, and communications

Sample Handling

  • StandardProvide a representative natural-gas sample within the selected pressure, temperature, and flow limits
  • Application DependentManage free liquids, glycol, methanol, compressor oil, aerosols, particulates, and heavier hydrocarbons without changing the property being measured
  • Application DependentDefine whether the measurement is made in a flow-through sample system or directly at the pipeline module and state the pressure basis

Installation & Utilities

  • StandardInstall the Dew Point Transducer in the approved hazardous area and the Central Control Unit in an explosion-safe area
  • Project EngineeredAnalyzer shelter, sample panel, cable route, power distribution, grounding, heating, ventilation, drainage, and sample return or vent

Integration & Communications

  • Project EngineeredDefine output scaling, register map, pressure and temperature inputs, correction logic, alarms, quality flags, historian tags, and DCS or SCADA ownership

Calibration & Quality Control

  • StandardFirst-principle chilled-mirror measurement and configuration-specific factory documentation support the supplied range and accuracy
  • Application DependentEstablish installed verification using an approved reference hygrometer, contractual procedure, and the relevant measurement pressure

Maintenance & Service

  • StandardUse the smart mirror-cleaning function and inspect the mirror, sample path, filters, seals, pressure-control items, and cooling hardware according to the manual
  • Application DependentMaintenance frequency depends on liquid or aerosol exposure, gas cleanliness, installation method, and the site quality plan

Safety & Application Considerations

  • StandardConfirm pressure rating, hazardous-area certification, safe-area separation, power supply, gas compatibility, isolation, depressurization, and sample disposal
  • Application DependentDirect-pipeline, high-pressure, and supplemental-cooling configurations require approved mechanical and electrical project design
Application Review

Application Engineering Notes

Range and cooling selection

Water and hydrocarbon measurement ranges are selected configurations rather than one continuous combined range. Supplemental cooling is required for water dew-point measurements below -30 °C.

Ordered configuration

The approved order identifies the selected ranges, required water-dew-point accuracy, measurement pressure, pressure and temperature inputs, sample-system or direct-pipeline arrangement, and Central Control Unit power supply.

Project Delivery

Project Integration & Support Scope

Blue Dragon reviews the measurement objective, interfaces, documentation, commissioning, and lifecycle requirements as part of project configuration. The scope below identifies the principal items to confirm before order placement.

Measurement Objective
Automatic online measurement of water dew point, hydrocarbon condensation temperature, or both alternately in natural gas, with safe-area operation and plant-data interfaces through the Central Control Unit.
Signal Outputs
The Central Control Unit provides RS-485 Modbus RTU, RS-232 Modbus RTU, and two 4-20 mA outputs. It can accept 4-20 mA pressure and temperature inputs. Final scaling, register mapping, correction logic, and alarm use are project specific.
Data Integration
Integrate the Central Control Unit with the DCS, PLC, SCADA, historian, or metering system. Define reported parameters, measurement mode, engineering units, pressure basis, pressure and temperature inputs, quality flags, and ownership of corrected values.
Alarms & Diagnostics
The project should define analyzer status and diagnostic handling, high water-dew-point and hydrocarbon-condensation alarms, sample-system alarms, cooling status, communication failure behavior, and local or control-room indication.
Documentation
The project package may include the manufacturer product sheet and operating manual, approved datasheet, metrological and hazardous-area certificates, dimensional and wiring information, communication details, sample-system drawings, and FAT or SAT records as ordered.
Commissioning Support
Commissioning can include mechanical and electrical inspection, pressure and leak testing, flow setup, cooling checks, mirror and sample-path inspection, Central Control Unit setup, analogue and Modbus tests, alarm simulation, and reference comparison.
Lifecycle Support
Lifecycle support covers mirror and sample-system maintenance, filters, seals, pressure and temperature inputs, cooling equipment, communications, verification, spare parts, metrological service coordination, troubleshooting, and technical assistance through Blue Dragon Technology.
Online Analyzer Configuration

Sample System & Online Analyzer Readiness

Reliable online measurement depends on the process interface as well as the analyzer. Confirm the following application data before final configuration and quotation.

Sample Type & Phase
Representative single-phase natural gas at the defined measurement pressure. Provide full composition and identify free liquids, glycol, methanol, compressor oil, aerosol, particulate, and expected heavy-hydrocarbon content.
Measurement Range
Select the required water and hydrocarbon ranges, water accuracy option, measurement mode, alarm points, measurement pressure, and any contractual reference-pressure correction. Identify whether supplemental cooling is required.
Sample Conditions
Provide pipeline pressure and temperature, intended measurement pressure, available flow, ambient range, gas composition, contaminant loading, safe-area distance, and the sample vent or return conditions.
Sample Handling
Define the gas-preparation system or direct-pipeline module, extraction probe, filtration philosophy, pressure and flow control, cooling, bypass, liquid handling, pressure and temperature inputs, calibration connection, and sample return or disposal.
Installation Area
Confirm hazardous-area classification and location for the transducer, safe-area location for the Central Control Unit, ambient exposure, shelter or panel space, cable route, direct-pipeline mechanical requirements, and maintenance access.
Utilities
Confirm the Central Control Unit and accessory power supplies, approved hazardous-area wiring, grounding, sample flow, supplemental-cooling utilities, heating, ventilation, drainage, and safe vent or return.
Response Requirement
State required cycles per hour, sample transport distance, acceptable stabilization time, alarm delay, and response expectations following process changes, range changes, or alternating water and hydrocarbon measurement.
Integration Requirement
Define RS-485 and RS-232 Modbus use, two 4-20 mA output assignments, pressure and temperature input scaling, correction logic, alarm mapping, quality flags, historian tags, and operator display requirements.
Commissioning Requirement
Confirm mechanical and electrical checks, pressure and leak tests, flow setup, cooling test, pressure and temperature input checks, output and Modbus tests, alarm simulation, reference comparison, training, and records.