- Sample Type & Phase
- Representative single-phase natural gas at the defined measurement pressure. Provide full composition and identify free liquids, glycol, methanol, compressor oil, aerosol, particulate, and expected heavy-hydrocarbon content.
- Measurement Range
- Select the required water and hydrocarbon ranges, water accuracy option, measurement mode, alarm points, measurement pressure, and any contractual reference-pressure correction. Identify whether supplemental cooling is required.
- Sample Conditions
- Provide pipeline pressure and temperature, intended measurement pressure, available flow, ambient range, gas composition, contaminant loading, safe-area distance, and the sample vent or return conditions.
- Sample Handling
- Define the gas-preparation system or direct-pipeline module, extraction probe, filtration philosophy, pressure and flow control, cooling, bypass, liquid handling, pressure and temperature inputs, calibration connection, and sample return or disposal.
- Installation Area
- Confirm hazardous-area classification and location for the transducer, safe-area location for the Central Control Unit, ambient exposure, shelter or panel space, cable route, direct-pipeline mechanical requirements, and maintenance access.
- Utilities
- Confirm the Central Control Unit and accessory power supplies, approved hazardous-area wiring, grounding, sample flow, supplemental-cooling utilities, heating, ventilation, drainage, and safe vent or return.
- Response Requirement
- State required cycles per hour, sample transport distance, acceptable stabilization time, alarm delay, and response expectations following process changes, range changes, or alternating water and hydrocarbon measurement.
- Integration Requirement
- Define RS-485 and RS-232 Modbus use, two 4-20 mA output assignments, pressure and temperature input scaling, correction logic, alarm mapping, quality flags, historian tags, and operator display requirements.
- Commissioning Requirement
- Confirm mechanical and electrical checks, pressure and leak tests, flow setup, cooling test, pressure and temperature input checks, output and Modbus tests, alarm simulation, reference comparison, training, and records.